Scope of work (Low Current System)
Provide a complete turnkey solution to install a fully operational Abbott laboratory: conduct a site assessment and design the lab layout to meet Abbott’s specifications, prepare the facility infrastructure (HVAC, electrical, plumbing, and data cabling), deliver, assemble, and install all Abbott instruments and any automation track system, perform installation and operational qualification (IQ/OQ), integrate with Abbott informatics, test end to end performance, and train laboratory staff before handing over the lab with full documentation and a service agreement.
1. Site Assessment & Design
Conduct an on-site survey to evaluate existing space, utilities (power, water, HVAC), and network infrastructure.
Determine the optimal layout for Abbott instruments, automation track, and workstations to ensure efficient workflow.
Prepare detailed engineering drawings and utility requirement plans based on Abbott’s technical specifications.
2. Infrastructure Preparation
Upgrade electrical supply – dedicated circuits, surge protection, and emergency power connections (UPS/generator) for all Abbott analyzers.
Modify the HVAC system to maintain the required temperature and humidity ranges for instrument operation.
Install a purified water system (Reverse Osmosis / Deionized) where required (e.g. for ARCHITECT or Alinity systems).
Run compressed air lines (oil-free, dried) where needed for pneumatic systems.
Install structured data cabling (Cat6/fiber) and network switches for LIS (Laboratory Information System) integration.
3. Equipment Delivery, Assembly & Installation
Coordinate delivery, unpacking, and positioning of all Abbott instruments (Alinity series, ARCHITECT, i-STAT, Cell-DYN, etc.).
Assemble and install each analyzer according to Abbott’s installation checklist.
Connect instruments to power, data, water, and compressed air.
Install Abbott automation track modules (where specified) and connect them to the analyzers.
Set up Abbott Informatics systems (e.g. AIM, STARLIMS) on dedicated servers/workstations.
4. Qualification, Integration & Testing
Perform Installation Qualification (IQ) and Operational Qualification (OQ) for every instrument per Abbott protocols.
Integrate all analyzers with the Laboratory Information System (LIS) and middleware.
Conduct end-to-end testing: sample loading, processing, results reporting, and data backup.
Simulate a power failure to verify seamless transition to backup power.
Test all safety and alarm systems.
5. Training & Documentation
Provide role-based training for lab supervisors, technicians, and IT staff (instrument operation, maintenance, software use, troubleshooting).
Deliver complete documentation: as-built drawings, IQ/OQ reports, user manuals, service contacts, and warranty certificates.
Schedule on-site post-installation support for 2–4 weeks (or as agreed) to ensure a smooth transition.